The specifications for the eFlo instrument are as follows.

 

Weight

10.4 lbs (4.76kg)

Power Required

24 VDC@ 750 mA

Enclosure Rating

IP10

Accuracy

4%

Repeatability

2%

Turndown Ratio

6:1

Medium Temperature Limits

-10°F to 125°F (-20°C to 51°C)

Ambient Temperature Limits

-10°F to 125°F (-20°C to 51°C)

Maximum Altitude

N/A

Flow Output Signal (Linear)

4-20 mA

Maximum Output Signal Load

500Ω

Input Control Signal (Linear)

4-20 mA

Communications

RS485, Ethernet, USB

Communication Protocol

Modbus RTU

 

 

Flow Meter Pressure Limits

 

The installed sensors will be determined by SSI based on pressure specifications & turndown requirements.

 

See the Calibration Report included with your product for more details.

2 psig sensors:

  • 2 psig usable limit

  • 4 psig overpressure limit

 

5 psig sensors:

  • 5 psig usable limit

  • 10 psig overpressure limit

 

Table 1 - eFlo Specifications