9.2.2 To Adjust Cold Junction Compensation (CJC)

 

Thermocouples are calibrated by following the previous procedure for the mV ranges (Section 9.2.1), then calibrating the CJC. Set ‘P1’ to J thermocouple type and verify that the temperature units of measurement is set for degrees Celsius. It is important to ensure P3 (Low Scale) & P4 (High Scale) ranges have been set for the J T/C type.

 

Connect a J T/C source as described in section 9.1.3. Set the mV source to ‘internal compensation’ for the thermocouple in use and set the output of the mV source to 0mV. If your sourcing device does not have this feature then set it to source 0.0 degrees Celsius to the instrument on the V+ and V- terminals.

 

Then:-

 

Operation

Do This

Display View

Additional Notes

Select calibration phase

1.        Press or to select ‘PHAS

This is found towards the end of the list of ‘P’ codes

Select CJC calibration

2.        Press or to select ‘CJC

 

Calibrate CJC

3.        Press to select ‘GO

4.        Press or to choose ‘YES

5.        Press to confirm

6.        Press again to confirm and to exit from the calibration phase

 

 

 

 

 

The controller automatically calibrates to the CJC input at 0mV.

The display will show busy then pass, (if calibration is successful) or ‘FAIL’ if not.

Fail may be due to an incorrect input mV