9.2.1 To Calibrate mV Input
NOTE: Set P code ‘P1’ to mV range (Nv). Ensure that the P3 (Low Scale) and P4 (High Scale) have been set for 0 to 80 mV. Copper wire will be needed when sourcing mV to the instrument.
Connect a 0 – 50mV source as shown in section 9.1.2. mA calibration is included in this procedure.
For best results 0mV should be calibrated by disconnecting the copper wires from the mV source and short circuiting the input to the controller.
Select Configuration Level as described in section 5.1.
Set P code ‘P1’ to mV range (Nv), then:-
Operation |
Do This |
Display View |
Additional Notes |
Select calibration phase |
1. Press or to select PHAS |
|
This is found towards the end of the list of ‘P’ codes. |
Set mV source for 0mV |
|||
Select the low calibration point |
2. Press or to choose ‘0’ |
|
|
Calibrate the instrument to the low calibration point (0mV) |
3. Press to select ‘GO’ 4. Press or to choose ‘YES’ 5. Press to confirm
|
|
The controller automatically calibrates to the injected input mV. The display will show busy for approximately 3-10 seconds, then pass, (if calibration is successful).
If FaiL is shown this is because the reference measurement is unstable or outside +/-20% of expected mV input.
If this should occur it will be necessary to go back to the beginning of the calibration, ensure that the correct mV are set and repeat the step that failed. |
|
6. Press again to enter |
|
|
Set mV source for 50mV |
|||
Select the high calibration point |
7. Press to go back to ‘PHAS’ 8. Press or to choose ‘50’ 9. Press to enter |
|
The controller will again automatically calibrate to the injected input mV.
If it is not successful then ‘FAIL’ will be displayed |
|
10. Press or to choose ‘YES’ 11. Press to confirm 12. Press again to confirm and exit from the calibration phase |
|
Note: PASS is displayed at the end of a calibration to indicate that the operation has been successful and has produced a value that appears to be within tolerance (+/- 20%) of expected values. It does not indicate that that the calibration is precisely correct, and the measured value must be verified in instrument operator mode against known sources to ensure calibration accuracy. |