9.1.1. Precautions
Before checking or starting any calibration procedure the following precautions should be taken:-
-
RTD and CJC calibration must not be carried out without prior mV calibration.
-
When calibrating mV inputs make sure that the calibrating source output is set to less than 250mV before connecting it to the mV terminals. If a large potential is accidentally applied (even for less than 1 second), then at least one hour should elapse before commencing the calibration.
-
A pre-wired jig built using a spare instrument sleeve may help to speed up the calibration procedure especially if a number of instruments are to be calibrated.
-
Power should be turned on only after the controller has been inserted in the sleeve of the pre-wired circuit. Power should also be turned off before removing the controller from its sleeve.
-
Ten minutes should be allowed for the controller to warm up after switch on.