Span Calibration Procedure
-
First tap “Zero/Span” and use the “Toggle Zero/Span” button to select Span.
-
Tap to highlight “Enter Gas Concentration” and tap the “Enter Gas %” button.
-
Enter the appropriate concentration of the calibration gas (see note on the previous page about the gas chemistry certification).
-
Begin the flow of gas and wait until the gas reading on the display stabilizes. This occurs when the actual values are not moving in a specific direction, and they display only slight movements up and down. This should take approximately 45 seconds.
-
Once stabilized, tap to highlight “Run Calibration” and tap the “Run Calibration” button.
-
The Calibration Timer on the screen will count down, and when it reaches zero the current gas value will adjust to match the gas concentration.