Output Configuration
The Output Configuration screen allows you to adjust output parameters for loops 1 and 2.
For each loop, the following parameters can be adjusted:
-
Source: A selected source: H2, DA, NH3, KN, External, or Standard KN.
-
Zero (%): Is the value that corresponds to 4 mA on a 4-20 mA scale. (4-20 mA is the default Range setting. If Range is set to 0-20 mA, then the Zero % refers to 0 mA.).
-
Span (%): Is the value that corresponds to 20 mA on a 4-20 mA scale. (4-20 mA is the default Range setting. If Range is set to 0-20 mA, then the Span % still refers to 20 mA.).
-
Range: Allows the user to choose between an output signal of 4-20 mA (default) and 0-20 mA.
-
Manual (%): Allows the user to enter an output value to test the analog output. In order for this option to function, Source must be set to External.
Use the applicable “Set” button to set each parameter (for example, use “Set Source” to set the source).