CJ Trim
The CJ Trim Calibration Type is a more traditional way to calibrate the Cold Junction temperature. To begin, note the difference between the displayed temperature (PV 1 or PV 2 value) and the sourced temperature. The difference between the two is the adjustment in corresponding CJ 1 or CJ 2 that will need to be entered.
-
Select the CJ Trim Calibration Type.
-
Select the correct input in the Inputs field.
-
Source a temperature to that input.
-
Note the difference between the sourced temperature and the value in the corresponding PV 1 or PV 2 field.
-
Add/Subtract that value to the corresponding CJ 1 or CJ 2 value indicated at the top of the page. For example, PV 2 is being sourced with 1000°F. PV 2 reads 998°F and CJ 2 reads 77.95°F. The difference between the sourced temperature and PV 2 is -2°F so we subtract 2°F to 77.95°F, which equals 75.95°F. This value is what will be entered in step 6.
-
Enter that new value into the Enter Target field and push the Calibrate button. A timer will count down from 30 seconds.
-
Once the Calibration Timer has counted down, the corresponding CJ 1 or CJ 2 value should match what was entered into the Enter Target field and the corresponding PV 1 or PV 2 value should matched the source temperature.
-
Compare the sourced temperature with the corresponding PV 1 or PV 2 value. Make further adjustments as-needed and perform additional calibrations.