Zero & Span Calibration:
-
Select the correct input type for Input 1 from the drop-down menu. Press the Set Type button to the right in order to set that type of input.
-
Go down to the Calibration Type option and choose the Zero VDC.
-
Select Input 1 for the Inputs option.
-
If the input is a thermocouple, select the correct TC Type. If the input is not a thermocouple, the value in this field does not matter.
-
Select the appropriate Range for the input type selected. For thermocouple inputs, see the table above for the corresponding range.
-
Connect the calibrator and source a zero input signal (0 mV or 0 VDC).
-
Allow the Raw VDC 1 signal to stabilize.
-
Enter the same zero value as what is being sourced from the calibrator into the Enter Target field and press the Calibrate button. A timer will count down from 30 seconds.
-
Once the Calibration Timer has counted down, the zero calibration is complete. You should see the Raw VDC 1 value change to match the calibrator.
NOTE: The remaining steps for the span input calibration will be very similar to the steps performed for the zero input calibration.
-
The Input type selected in step 1 remains the same. Change the Calibration Type to Span VDC.
-
The TC Type and Range selected in steps 4 and 5 remain the same.
-
Using the calibrator, source a span input signal. The span signal should be at least 90% of the full range of the sensor (for example, source at least 1.125 VDC for a sensor with a range of 0 to 1.25 VDC).
-
Allow the Raw VDC 1 signal to stabilize.
-
Enter the same span value as what is being sourced from the calibrator into the Enter Target field and press the Calibrate button. A timer will count down from 30 seconds.
-
Once the Calibration Timer has counted down, the span calibration is complete. You should see the Raw VDC 1 value change to match the calibrator.
This same procedure can be performed for Input 2 if it is being used. You just need to select the correct Input type in the Input 2 field and choose the Input 2 for the Inputs field.