Specifications
Gas Measurement Specifications
|
CO2 Sensor |
|
|
Standard Range |
0 – 2.000% |
|
Optional High Range |
0 – 20.00% |
|
Accuracy (Standard Range) |
±0.006% |
|
Accuracy (High Range) |
±0.2% |
|
Resolution (Standard Range) |
±0.001% |
|
Resolution (High Range) |
±0.01% |
|
Measurement Method |
Non-Dispersive Infrared (NDIR) |
|
CO Sensor |
|
|
Range |
0 – 100.00% |
|
Accuracy |
±0.2% |
|
Resolution |
±0.01% |
|
Measurement Method |
Non-Dispersive Infrared (NDIR) |
|
CH4 Sensor |
|
|
Range |
0 – 100.00% |
|
Accuracy |
±0.2% |
|
Resolution |
±0.01% |
|
Measurement Method |
Non-Dispersive Infrared (NDIR) |
Single Gas Analyzer (SGA)
|
Response Time |
0 – 6 seconds |
|
Power Supply Input Voltage |
110VAC or 230VAC |
|
Maximum Operating Temperature |
122 °F (50 °C) |
|
Analog Outputs |
2 (4-20mA or 0-5 V) |
|
Serial Communications |
2 RS485 ports using Modbus RTU, configurable baud rate |
|
Ethernet |
1 port |
|
USB |
1 Type A port, 1 Type B port |