Specifications

 

Gas Measurement Specifications

 

CO2 Sensor

Standard Range

0 – 2.000%

Optional High Range

0 – 20.00%

Accuracy (Standard Range)

±0.006%

Accuracy (High Range)

±0.2%

Resolution (Standard Range)

±0.001%

Resolution (High Range)

±0.01%

Measurement Method

Non-Dispersive Infrared (NDIR)

CO Sensor

Range

0 – 100.00%

Accuracy

±0.2%

Resolution

±0.01%

Measurement Method

Non-Dispersive Infrared (NDIR)

CH4 Sensor

Range

0 – 100.00%

Accuracy

±0.2%

Resolution

±0.01%

Measurement Method

Non-Dispersive Infrared (NDIR)

 

Single Gas Analyzer (SGA)

 

Response Time

0 – 6 seconds

Power Supply Input Voltage

110VAC or 230VAC

Maximum Operating Temperature

122 °F (50 °C)

Analog Outputs

2 (4-20mA or 0-5 V)

Serial Communications

2 RS485 ports using Modbus RTU, configurable baud rate

Ethernet

1 port

USB

1 Type A port, 1 Type B port