Product Specifications
-
User defined “logging” intervals as frequently as 10 seconds
-
Isolated inputs
-
AC ( 90 – 240 V) and DC Power
-
Rechargeable battery – Approximately 12 hours of “runtime”
-
Dimensions: 16”D X 20”L X 8”H
-
Password protected menu options
-
Easy calibration
-
Web-page enabled real-time view
-
ETHERNET and RS232 serial communications
-
16-Bit A/D converter
-
128 Mbytes of on-board data
|
Degrees F. |
A/D |
Accuracy |
Accuracy |
Accuracy |
|
TC Type |
Min |
Max |
Full scale |
Full Scale |
At 1600 Degrees F. |
At 1600 Degrees F with thermister error |
B |
32 |
3308 |
0.0124% |
0.0179% |
0.29 |
0.65 |
C |
32 |
4208 |
0.0062% |
0.0067% |
0.11 |
0.47 |
E |
-328 |
1832 |
0.0031% |
0.0032% |
0.05 |
0.41 |
J |
-346 |
2192 |
0.0031% |
0.0036% |
0.06 |
0.42 |
K |
-328 |
2502 |
0.0031% |
0.0045% |
0.07 |
0.43 |
N |
-328 |
2372 |
0.0031% |
0.0052% |
0.08 |
0.44 |
NNM |
0 |
1409 |
0.0031% |
0.0033% |
0.05 |
0.41 |
R |
-58 |
3214 |
0.0062% |
0.0118% |
0.19 |
0.66 |
S |
-58 |
3214 |
0.0124% |
0.0133% |
0.21 |
0.57 |
T |
-328 |
752 |
0.0062% |
0.0119% |
0.19 |
0.55 |
*** Note: Accuracy at 1600 Degrees F with thermister error requires SDS firmware level 1.10 or later ***