Appendix 1: PGA 3510 Common Modbus Registers
This appendix covers Modbus registers most commonly accessed by PGA 3510 users. These are Modbus registers 200 through 227 and their descriptions. There are other Modbus registers used by the PGA 3510. If you need information on other registers, call SSi at (513) 772-0060.
Register Location |
Parameter |
Description |
---|---|---|
200 |
co_data |
Measured CO value |
201 |
co2_data |
Measured CO2 value |
202 |
o2_data |
Measured 02 value |
203 |
ch4_data |
Measured CH4 value |
204 |
Bench_off |
Turn off bench, pump, and sample |
205 |
IR_carbon |
IR %C |
206 |
IR_COF |
Suggested COF (CO Factor) |
207 |
IR_PF |
Suggested PF (Process Factor) |
208 |
IR_MV |
IR equivalent millivolts |
209 |
probe_MV |
Probe millivolts |
210 |
probe_COF |
Probe COF |
211 |
probe_temperature |
Probe temperature |
212 |
probe_PF |
Probe PF |
213 |
probe_carbon |
Probe %C |
214 |
pump_run_min |
The total time in minutes that the pump has run since the last time memory was cleared or a manual reset took place. This value, plus pump_run_wks, will provide the total amount of time the pump has been running. When this number reaches 10080, pump_run_wks is increased by 1, and pump_run_min is reset to zero. |
215 |
pump_run_wks |
The total time in weeks that the pump has run since the last time memory was cleared or a manual reset took place. This value, plus pump_run_min, will provide the total amount of time the pump has been running. |
216 |
pump |
Pump status |
217 |
pump_run_now |
The time in minutes that the pump is running (reset to zero if pump is turned off). Maximum value is 30000. If pump_run_now is greater than pump_run_max, then pump_run_max is set to pump_run_now. |
218 |
pump_run_max |
The longest time in minutes of a pump session. If pump_run_now is greater than pump_run_max, then pump_run_max is set to pump_run_now. |
219 |
calib_status |
Calibration status |
220 |
IR_temperature |
IR temperature |
221 |
output_percent |
Control output in percent |
222 |
sample_flow |
Sample flow |
223 |
sample_valve |
Sample valve state |
224 |
sample_control |
Sample valve control |
225 |
flow_timer |
Flow control |
226 |
base_status |
Status of base (0=OFF, 1=pump ON, 2=sample delay, 3=measure delay, 4=delay OFF, 5=ON) |
227 |
alarm_bits |
Alarm bit map (Bit 0=low flow, Bit 1=base communications, Bit 2=max factor, Bit 3=min factor, Bit 6=programmer alarm, Big 7=high byte is acknowledge; Bit 4 and 5 not used) |