Appendix 1: PGA 3510 Common Modbus Registers

 

This appendix covers Modbus registers most commonly accessed by PGA 3510 users. These are Modbus registers 200 through 227 and their descriptions. There are other Modbus registers used by the PGA 3510. If you need information on other registers, call SSi at (513) 772-0060.

 

Register Location

Parameter

Description

200

co_data

Measured CO value

201

co2_data

Measured CO2 value

202

o2_data

Measured 02 value

203

ch4_data

Measured CH4 value

204

Bench_off

Turn off bench, pump, and sample

205

IR_carbon

IR %C

206

IR_COF

Suggested COF (CO Factor)

207

IR_PF

Suggested PF (Process Factor)

208

IR_MV

IR equivalent millivolts

209

probe_MV

Probe millivolts

210

probe_COF

Probe COF

211

probe_temperature

Probe temperature

212

probe_PF

Probe PF

213

probe_carbon

Probe %C

214

pump_run_min

The total time in minutes that the pump has run since the last time memory was cleared or a manual reset took place. This value, plus pump_run_wks, will provide the total amount of time the pump has been running. When this number reaches 10080, pump_run_wks is increased by 1, and pump_run_min is reset to zero.

215

pump_run_wks

The total time in weeks that the pump has run since the last time memory was cleared or a manual reset took place. This value, plus pump_run_min, will provide the total amount of time the pump has been running.

216

pump

Pump status

217

pump_run_now

The time in minutes that the pump is running (reset to zero if pump is turned off). Maximum value is 30000. If pump_run_now is greater than pump_run_max, then pump_run_max is set to pump_run_now.

218

pump_run_max

The longest time in minutes of a pump session. If pump_run_now is greater than pump_run_max, then pump_run_max is set to pump_run_now.

219

calib_status

Calibration status

220

IR_temperature

IR temperature

221

output_percent

Control output in percent

222

sample_flow

Sample flow

223

sample_valve

Sample valve state

224

sample_control

Sample valve control

225

flow_timer

Flow control

226

base_status

Status of base (0=OFF, 1=pump ON, 2=sample delay, 3=measure delay, 4=delay OFF, 5=ON)

227

alarm_bits

Alarm bit map (Bit 0=low flow, Bit 1=base communications, Bit 2=max factor, Bit 3=min factor, Bit 6=programmer alarm, Big 7=high byte is acknowledge; Bit 4 and 5 not used)