Specifications

The unit is designed and manufactured for the atmosphere heat-treating industry.

 

CO:

Range: 0.00 to 30.00 % of gas concentration

Accuracy: +/-1% of full scale (+/-0.3% of gas concentration, based on 30%)

Resolution: 0.01%

 

CO2:

Low range: 0.000 to 2.000 % of gas concentration

Accuracy: +/-1% of full scale (+/-0.02% of gas concentration, based on 2%)

Resolution: 0.001%

 

CO2:

High range: 0.00 to 20.00% of gas concentration

Accuracy: +/-1% of full scale (+/-0.2% of gas concentration, based on 20%)

Resolution: 0.01%

 

CH4:

Range: 0.00 to 20.00 % of gas concentration

Accuracy: +/-1% of full scale (+/-0.2% of gas concentration, based on 20%)

Resolution: 0.01%

 

O2:

Range: 0.01 to 25.00% of gas concentration

Accuracy: +/-1% of full scale

 

* Note: The CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown above, however both sensors are capable of measuring gas concentrations of up to 100.00%.

 

Sampling method: Extraction by internal pump (when necessary)
Accuracy and repeatability:  ± 1% of full scale
Flow Meter:  Inside case lid and also on-screen
Pump Operation: On/Off/Automatic
AC Power: 90 to 230 VAC, 50 to 60 Hz, 60 Watts
DC Power:

12VDC rechargeable NiCd battery (Charger integrated inside instrument)

Communications:  RS232, Ethernet, RS485 (Future Use)  
Data Storage:  Continuous automatic data logging
Data Retrieval:

Visible in graphical and tabular format using XGA Viewer software (included)

Operating Temperature:  32° to 122° F (0° to 50° C)
Dimensions: Approx. 16”H X 20”L X 8”D
Weight: Approx. 30 lbs.