Variable Output Filtering (Td Averaging)
The Mini-OX is factory default to use adaptive output filtering to give an optimum balance between output stability and response to oxygen changes. This balance may be altered to suit the needs of your application.
-
Connect the Mini-OX via the RS232 interface to the PC.
-
Press <ENTER>; the Menu Password screen displays.
-
Input your security password.
-
Press <ENTER> to access the Menu screens.
-
Type “2” to access theConfiguration menu.
-
Type “2” to access theEnter Td Averaging screen.
-
Input the required number, between 0 and 200; 0 for adaptive filtering (recommended), 1 for very fast and dynamic output response but relatively unstable to 200 for an extremely stable output but very slow response to oxygen changes.
-
Press <ENTER> to save.
NOTE: The new averaging value is now stored in memory and is retained on power loss.