Automatic Calibration Value
The system is factory set to automatically calibrate to 20.7% O2 to allow simple calibration in fresh air. The auto calibration value is factory set to 20.7% to take into account average humidity in the atmosphere. If a calibration with a gas of a different known O2 concentration is required, the factory set value may be changed via the RS232 interface.
-
Connect the Mini-OX via the RS232 interface to the PC.
-
Press <ENTER>; the Menu Password screen displays.
-
Input your security password.
-
Press <ENTER> to access the Menu screens.
-
Type “2” to access theConfiguration menu.
-
Type “1” to access the Enter Auto Calibscreen.
-
Input the oxygen concentration (%) of the calibration gas as a number to 2 decimal places.
-
Press <ENTER> to save.
NOTE:The new automatic calibration value is now stored in memory and is retained on power loss.
NOTE:If calibration is required with a different gas of known O2 concentration and access to the RS232 menus via a PC is not available, a manual calibration must be performed. Refer to Manual Calibration.