Analog Output Minimum and Maximum Ranges
The Mini-OX is factory default to output a range of 0 – 25% O2 via its two analog outputs. This range can be expanded to 0 – 100% O2 by repositioning the jumper link as described in Figure 7. When the unit is reconfigured to output 0 – 100% O2 the user also has the option to fully customize the output ranges via RS232. This is extremely useful in applications where the O2 variation is within a narrow band as it allows the analogue outputs to be tailored to this limited range.
NOTE: The minimum and maximum range adjustment does NOT apply to the RS232 output and is overruled if the unit is reconfigured for 0 – 25% O2 operation.
-
Ensure the Mini-OX is configured for 0 – 100% O2 operation, see Figure 7.
-
Connect the Mini-OX via the RS232 interface to the PC.
-
Press <ENTER>; the Menu Password screen displays.
-
Input your security password.
-
Press <ENTER> to access the Menu screen.
-
Type “2” to access theConfiguration menu.
-
Type “3” to access theEnter O2 Max Range screen.
-
Input the required number, between 1.00 and 100.00 to represent the maximum output range.
NOTE:The number must also be greater than the saved minimum range.
-
Press <ENTER> to save.
-
Press <ESC> to return to the Configuration menu.
-
Type “4” to access theEnter O2 Min Range screen.
-
Input the required number, between 0.00 and 99.00 to represent the minimum output range.
NOTE:The number must also be less than the saved maximum range.
-
Press <ENTER> to save.
NOTE:The new ranges are now stored in memory and are retained on power loss.
An example of changing the minimum and maximum output ranges would be in a fresh air atmosphere where the O2 range is between 20-21%. The user could set the minimum output range to 19% and the maximum output range to 22% and the outputs would vary linearly between. The minimum and maximum ranges lock out the outputs at the set limits so 19% O2 or lower would set the analog outputs to 0VDC/4mA and 22% O2or higher would set the analog outputs to 10VDC/20mA.