Analog Variants – 0-10VDCand 4-20mA Output Values
O2% |
Output Values |
|||
0 – 10VDC output |
4 – 20mA output |
|||
0.1 – 25% O2 |
0.1 – 100% O2 |
0.1 – 25% O2 |
0.1 – 100% O2 |
|
20.7% |
8.28VDC |
2.07VDC |
17.25mA |
7.34mA |
100% |
- |
10VDC |
- |
20mA |
90% |
- |
9.0VDC |
- |
18.4mA |
25% |
10VDC |
2.5VDC |
20mA |
8mA |
5% |
2.0VDC |
0.5VDC |
7.2mA |
4.8mA |
0.01% (see Note) |
0.04VDC |
0.01VDC |
4.06mA |
4.02mA |
NOTE: The analog output ranges actually represent 0 to 25% or 0 to 100% O2 however as SSi’s oxygen sensors cannot measure below 0.1% O2 this value is displayed as the range minimum.