APPENDIX B – Automatic Calibration using a Series 8 Controller (with Wiring Diagram and Configuration Information)
-
Place the sensor probe in the calibration gas, typically fresh air.
-
Press the F2 button on the Series 8.
-
Press the up arrow until the display reads “run,” then press the Enter button.
-
Wait about 30 seconds. When the display no longer reads “run,” calibration is complete.
For additional help, see Automatic Calibration section, Series 8 Manual, or contact SSi at 513-772-0060.
Configuration Settings |
|||
Series 804 - Mini-OX |
|||
Run Parameters |
Configuration |
||
Name |
Value |
Parameter |
Default |
OP |
0 |
P1 |
MV |
Unit |
NONE |
P2 |
NNN.N |
SPLo |
0.5 |
P3 |
0 |
SPHi |
20.9 |
P4 |
25 |
SP1 |
20.9 |
P5 |
0 |
SP2 |
0.5 |
P6 |
79.96 |
SPrr |
OFF |
P7 |
HP |
ALtSPLo |
-17.8 |
P11 |
NONE |
ALtSPHi |
204.4 |
P12 |
T.RUN |
SPTrim |
0 |
P13 |
HEAT |
Pb |
11.1 |
P14 |
NONE |
Ti |
360 |
P15 |
4.2 |
Td |
OFF |
P21 |
NONE |
cbLo |
AUTO |
P24 |
NONE |
cbHi |
AUTO |
P27 |
NONE |
Mr |
0 |
P31 |
NONE |
oFS |
0 |
P32 |
100 |
FiLt |
8 |
P33 |
NONE |
oPLo |
0 |
P34 |
OFF |
oPHi |
100 |
P35 |
OFF |
LdAL |
OFF |
P36 |
0 |
LEAL |
OFF |
P41 |
DLL |
HcAL |
OFF |
P42 |
MIN |
tdUr |
30s |
P43 |
RES |
tthr |
OFF |
P51 |
NONE |
|
|
P52 |
NONE |
|
|
P61 |
1 |
|
|
P62 |
19.2 |
|
|
P63 |
NONE |
|
|
P64 |
NONE |
|
|
P71 |
T.ST |
|
|
P72 |
T.ST |
|
|
P73 |
AC.AL |
|
|
P74 |
STD |
|
|
P75 |
OP |
|
|
P76 |
2 |
|
|
P77 |
4 |
|
|
P81 |
NONE |
|
|
rEcS |
NONE |
|
|
rEcL |
NONE |
|
|
PHAS |
NONE |