APPENDIX B – Automatic Calibration using a Series 8 Controller (with Wiring Diagram and Configuration Information)

  1. Place the sensor probe in the calibration gas, typically fresh air.

  2. Press the F2 F2jpg button on the Series 8.

  3. Press the up arrow Description: Up button.jpguntil the display reads “run,” then press the Enter Description: Scroll button.jpg button.

  4. Wait about 30 seconds.  When the display no longer reads “run,” calibration is complete.


For additional help, see Automatic Calibration section, Series 8 Manual, or contact SSi at 513-772-0060.

 

 

Configuration Settings

Series 804 - Mini-OX

Run Parameters

Configuration

Name

Value

Parameter

Default

OP

0

P1

MV

Unit

NONE

P2

NNN.N

SPLo

0.5

P3

0

SPHi

20.9

P4

25

SP1

20.9

P5

0

SP2

0.5

P6

79.96

SPrr

OFF

P7

HP

ALtSPLo

-17.8

P11

NONE

ALtSPHi

204.4

P12

T.RUN

SPTrim

0

P13

HEAT

Pb

11.1

P14

NONE

Ti

360

P15

4.2

Td

OFF

P21

NONE

cbLo

AUTO

P24

NONE

cbHi

AUTO

P27

NONE

Mr

0

P31

NONE

oFS

0

P32

100

FiLt

8

P33

NONE

oPLo

0

P34

OFF

oPHi

100

P35

OFF

LdAL

OFF

P36

0

LEAL

OFF

P41

DLL

HcAL

OFF

P42

MIN

tdUr

30s

P43

RES

tthr

OFF

P51

NONE

 

 

P52

NONE

 

 

P61

1

 

 

P62

19.2

 

 

P63

NONE

 

 

P64

NONE

 

 

P71

T.ST

 

 

P72

T.ST

 

 

P73

AC.AL

 

 

P74

STD

 

 

P75

OP

 

 

P76

2

 

 

P77

4

 

 

P81

NONE

 

 

rEcS

NONE

 

 

rEcL

NONE

 

 

PHAS

NONE