Appendix B: Inputs/Outputs on the MZA 6010

 

There are a number of Digital Inputs and Outputs in both the IR/interface enclosure and the pump enclosure on the MZA 6010.  These are usually tied to external events or activated by an external PLC to perform various functions.  The digital inputs are enabled by completing the electrical connection between each input and the ground terminal.

 

Digital I/O Board #1 in Main Enclosure

Input 1

Zone 1 COF/PF Inhibit

Input 2

Zone 2 COF/PF Inhibit

Input 3

Zone 3 COF/PF Inhibit

Input 4

Zone 4 COF/PF Inhibit

Output 1

Not used

Output 2

Alarm 2

Output 3

Alarm 1

Output 4

Aux. Cal Gas Enable

Output 5

Sample Gas Enable

Output 6

Autocal Zero Gas Enable

Output 7

Autocal Span Gas Enable

Output 8

Not Used

 

Digital I/O Board in Zone 1-4 Pump Enclosure

Input 1

Zone 1 Stop Sample

Input 2

Zone 2 Stop Sample

Input 3

Zone 3 Stop Sample

Input 4

Zone 4 Stop Sample

Output 1

Zone 1 Pump On/Off

Output 2

Zone 2 Pump On/Off

Output 3

Zone 3 Pump On/Off

Output 4

Zone 4 Pump On/Off

Output 5

Zone 1 Blocking Solenoid

Output 6

Zone 2 Blocking Solenoid

Output 7

Zone 3 Blocking Solenoid

Output 8

Zone 4 Blocking Solenoid

 

 

Digital I/O Board #2 in Main Enclosure

Input 1

Zone 5 COF/PF Inhibit

Input 2

Zone 6 COF/PF Inhibit

Input 3

Zone 7 COF/PF Inhibit

Input 4

Zone 8 COF/PF Inhibit

Output 1

Zone 1 Alarm

Output 2

Zone 2 Alarm

Output 3

Zone 3 Alarm

Output 4

Zone 4 Alarm

Output 5

Zone 5 Alarm

Output 6

Zone 6 Alarm

Output 7

Zone 7 Alarm

Output 8

Zone 8 Alarm

 

Digital I/O Board in Zone 1-4 Pump Enclosure

Input 1

Zone 5 Stop Sample

Input 2

Zone 6 Stop Sample

Input 3

Zone 7 Stop Sample

Input 4

Zone 8 Stop Sample

Output 1

Zone 5 Pump On/Off

Output 2

Zone 6 Pump On/Off

Output 3

Zone 7 Pump On/Off

Output 4

Zone 8 Pump On/Off

Output 5

Zone 5 Blocking Solenoid

Output 6

Zone 6 Blocking Solenoid

Output 7

Zone 7 Blocking Solenoid

Output 8

Zone 8 Blocking Solenoid

 

Shaded areas indicate internal functions of the MZA that are not available.  They are included here for reference and troubleshooting purposes.

 

 

 

The MZA 6010 also includes four (4) or eight (8) Analog Output Boards, depending on the model.  Each board corresponds to the same-numbered zone.  The outputs function as follows by default, but can be changed in the Analog Output Setup section on page 31.

 

Analog Board (#)

Output 1

Zone (#) %CO

Output 2

Zone (#) %CO2

Output 3

Zone (#) %CH4

Output 4

Zone (#) Carbon Potential (%C)