Appendix B: Inputs/Outputs on the MZA 6010
There are a number of Digital Inputs and Outputs in both the IR/interface enclosure and the pump enclosure on the MZA 6010. These are usually tied to external events or activated by an external PLC to perform various functions. The digital inputs are enabled by completing the electrical connection between each input and the ground terminal.
Digital I/O Board #1 in Main Enclosure |
|
Input 1 |
Zone 1 COF/PF Inhibit |
Input 2 |
Zone 2 COF/PF Inhibit |
Input 3 |
Zone 3 COF/PF Inhibit |
Input 4 |
Zone 4 COF/PF Inhibit |
Output 1 |
Not used |
Output 2 |
Alarm 2 |
Output 3 |
Alarm 1 |
Output 4 |
Aux. Cal Gas Enable |
Output 5 |
Sample Gas Enable |
Output 6 |
Autocal Zero Gas Enable |
Output 7 |
Autocal Span Gas Enable |
Output 8 |
Not Used |
Digital I/O Board in Zone 1-4 Pump Enclosure |
|
Input 1 |
Zone 1 Stop Sample |
Input 2 |
Zone 2 Stop Sample |
Input 3 |
Zone 3 Stop Sample |
Input 4 |
Zone 4 Stop Sample |
Output 1 |
Zone 1 Pump On/Off |
Output 2 |
Zone 2 Pump On/Off |
Output 3 |
Zone 3 Pump On/Off |
Output 4 |
Zone 4 Pump On/Off |
Output 5 |
Zone 1 Blocking Solenoid |
Output 6 |
Zone 2 Blocking Solenoid |
Output 7 |
Zone 3 Blocking Solenoid |
Output 8 |
Zone 4 Blocking Solenoid |
Digital I/O Board #2 in Main Enclosure |
|
Input 1 |
Zone 5 COF/PF Inhibit |
Input 2 |
Zone 6 COF/PF Inhibit |
Input 3 |
Zone 7 COF/PF Inhibit |
Input 4 |
Zone 8 COF/PF Inhibit |
Output 1 |
Zone 1 Alarm |
Output 2 |
Zone 2 Alarm |
Output 3 |
Zone 3 Alarm |
Output 4 |
Zone 4 Alarm |
Output 5 |
Zone 5 Alarm |
Output 6 |
Zone 6 Alarm |
Output 7 |
Zone 7 Alarm |
Output 8 |
Zone 8 Alarm |
Digital I/O Board in Zone 1-4 Pump Enclosure |
|
Input 1 |
Zone 5 Stop Sample |
Input 2 |
Zone 6 Stop Sample |
Input 3 |
Zone 7 Stop Sample |
Input 4 |
Zone 8 Stop Sample |
Output 1 |
Zone 5 Pump On/Off |
Output 2 |
Zone 6 Pump On/Off |
Output 3 |
Zone 7 Pump On/Off |
Output 4 |
Zone 8 Pump On/Off |
Output 5 |
Zone 5 Blocking Solenoid |
Output 6 |
Zone 6 Blocking Solenoid |
Output 7 |
Zone 7 Blocking Solenoid |
Output 8 |
Zone 8 Blocking Solenoid |
Shaded areas indicate internal functions of the MZA that are not available. They are included here for reference and troubleshooting purposes.
The MZA 6010 also includes four (4) or eight (8) Analog Output Boards, depending on the model. Each board corresponds to the same-numbered zone. The outputs function as follows by default, but can be changed in the Analog Output Setup section on page 31.
Analog Board (#) |
|
Output 1 |
Zone (#) %CO |
Output 2 |
Zone (#) %CO2 |
Output 3 |
Zone (#) %CH4 |
Output 4 |
Zone (#) Carbon Potential (%C) |