Mechanical Installation - Enclosure Mounting

It is recommended that the MGA 6010 be mounted as close to the sampling point as possible, since that will reduce the length of the plumbing lines that will need to be maintained.  It is intended for use in a heat treating environment, but care should be taken not to mount it too close to a furnace or other heat source.  The operating temperature of the enclosure should be maintained below 122°F (50°C).  If necessary, a heat shield can be mounted behind the enclosure to reduce the amount of radiant heat that the MGA is exposed to.  In most cases, this will not be necessary.

 

The enclosure is heavily vented to prevent the buildup of potentially harmful gases in the unlikely event of an internal leak.  This venting will also reduce the internal temperature by allowing the free flow of ambient air around the internal components.