Specifications

 

The unit is designed and manufactured for the atmosphere heat treating industry; however, its uses go beyond the scope of these applications.

 

CO:

Range: 0 – 30% of gas concentration.

Accuracy: +/-1% of full scale (+/-0.3% of gas concentration, based on 30%).

Resolution: 0.01%

 

CO2 Low:

Range: 0 – 2% of gas concentration.

Accuracy: +/-1% of full scale (+/-0.02% of gas concentration, based on 2%).

Resolution: 0.001%

 

CO2 High:

Range: 0 – 20% of gas concentration.

Accuracy: +/-1% of full scale (+/-0.2% of gas concentration, based on 20%).

Resolution: 0.01%

 

CH4:

Range: 0 – 20% of gas concentration.

Accuracy: +/-1% of full scale (+/-0.2% of gas concentration, based on 20%).

Resolution: 0.01%

 

H2:

Range: 0 – 100% of gas concentration.

Accuracy: +/-1% of full scale (+/-1% of gas concentration, based on 100%).

Resolution: 0.1%

 

* Note: These sensors have been optimized for use at the levels normally seen in an Endothermic atmosphere.  The ranges can be adjusted to fit other applications.  For information regarding modifications to the ranges shown above, please contact Super Systems.

 

Sampling method:   Extraction by internal pump (when necessary)
Measurement Method (CO, CO2, CH4): Non-Dispersive Infrared (NDIR)
Measurement Method (H2): Thermal Conductivity
Accuracy and repeatability: ± 1% of full scale
Recommended Calibration Interval: Annual
AC Power Requirements: 110VAC (can be modified to 220V upon request)
Communications: Ethernet, USB(A), USB(B), RS485 Modbus
Data Storage: Continuous automatic data logging
Data Retrieval:  XGA Viewer Software (included) or on-screen
Operating Temperature:  32° to 122° F (0° to 50° C)
External Dimensions: Approx. 16”H X 20”L X 8”D
Weight: Approx. 28 lbs.