Introduction

The Model MGA 6010 (see part numbers in the Parts List on page 1) is a Multi-Gas IR analyzer. It measures Carbon Monoxide (CO), Carbon Dioxide (CO2) and Natural Gas (CH4) typically found in an endothermic atmosphere.  The measurement of these gases, combined with furnace temperature information, allows the MGA 6010 to calculate the percent Carbon (%C) of the measured gas. A Hydrogen (H2) sensor can also be incorporated as an option to provide a more complete picture of the measured gas.

 

For nitriding and ferritic nitrocarburizing (FNC) applications, the MGA 6010 can calculate carburizing potential (KC) and nitriding potential (KN).

 

Finally, the MGA 6010 can be configured for compatibility with environments containing ammonia (NH3) gas. NH3 compatibility must be specifically requested when the MGA 6010 is ordered.

 

NOTE: Calculation of Kc may have inaccuracies in the presence of ammonia.

 

IMPORTANT!

 

Do not use a MGA 6010 for gas analysis with NH3-containing gas if the MGA 6010 has not been configured with NH3 compatibility. Any use of a non-NH3 compatible MGA 6010 with NH3-containing gas will void the product warranty.

 

 

 

Basic Operating Description

The Model MGA 6010 has been designed for the simultaneous analysis of CO, CO2 and CH4 in heat-treat furnace atmosphere gases. It uses a color touch screen display / operator interface for data entry and for viewing.  Selections can be made on the screen using a finger or a stylus.  Avoid the use of sharp objects (pens, paperclips, screwdrivers, etc.) as they can cause permanent damage to the screen and void the warranty of the instrument.

 

After the power switch is turned on, it will take approximately 30 seconds for the MGA 6010 software to automatically load.  Once the software is properly loaded, the instrument is ready to use.  Power to the MGA can be turned off by tripping the circuit breaker inside the enclosure. Before the breaker is tripped, select the “Shut down interface” option from the main menu and allow the screen to close the MGA software in a controlled manner. Doing so will help minimize any damage to data that could be done by an unexpected shutdown.

 

Hydrogen Cell Note:

For highest H2 accuracy, it is recommended that the instrument be powered on for 60 minutes before measurements are taken.

 

 

Dimensions: