General Specifications:
Total Measurement Range: |
-148°F to +32°F (-100°C to 0°C) |
Optimal Measurement Range: |
-112°F to -22°F (-80°C to -30°C) |
Accuracy (Within Optimal Range): |
+/- 3.6°F (+/- 2°C) |
Accuracy (Outside Optimal Range): |
+/- 5.4°F (+/- 3°C) |
Operating Temperature Range: |
-40°F to +140°F (-40°C to +60°C) |
Operating Pressure Range: |
0-725 psi / 0-50 bar |
Operating Humidity: |
Non-Condensing |
Sensor Housing Classification: |
IP65 – NEMA4 (AISI316L) |
Display Housing Classification: |
IP54 |
Display Weight: |
400 grams |
Power: |
Rechargeable NIMH Battery (included) or 4 “AA” Batteries |
Battery Recharging Time: |
Approximately 4 hours |
Battery Charger Input Requirements: |
110-240 VAC/ 47-63Hz / 400mA |
Run Time on Battery Power: |
48 Hours – Continuous use / Up to 30 days - Data logging only |
Output Units: |
Dew Point /Frost Point, ppm & ppb (by volume), ppm (by weight) |
Languages Supported: |
Nine Possible Languages (See “Language” section) |
Data Logging Capacity: |
2700 points, user-selectable measurement interval |
Data Logging Interval Options: |
1s, 5s, 15s, 30s, 1min, 5min, 15min, 30min, 1h, 3h, 12h |
Logging Duration Options: |
1min, 5min, 15min, 30min, 1h, 3h, 12h, 24h, 7d, 31d, Memory Full |