WARNINGS –

 

Although it is intended for use in an industrial environment, the DPC2530 is a sensitive piece of analysis equipment.  Care should be taken not to operate it in a manner inconsistent with its intended use. 

  • Moisture (water) cannot be allowed to enter the analyzer.  If water is present in the sample gas, use an in-line dryer for sample conditioning.  In the event that the sensor becomes wet, use an inert gas (such as nitrogen or argon) or dry compressed air to dry the inside of the instrument.  Under no circumstances should methane (or natural gas) be used to dry the DPC2530.

  • The analyzer must be stored at ambient temperature (65-80°F) for at least four hours prior to operation.

  • An in-line dryer for sample conditioning should be used for exothermic and combustion applications.

  • This unit is not designed to measure the dew points in corrosive gasses, such as ammonia (NH3), sulphur trioxide (SO3), chlorine (Cl), and hydrochloric acid (HCl).

  • Please read and understand this Operations Manual before operating the unit.

 

Failure to comply with these conditions may cause damage to the unit that will not be covered under the warranty.  Super Systems, Inc. is not responsible for damage to this unit caused by disregard of these warnings, neglect, or misuse.