WARNINGS –
Although it is intended for use in an industrial environment, the DP2000 is a sensitive piece of analysis equipment. Care should be taken not to drop the analyzer or to operate it in a manner inconsistent with its intended use.
-
Open all sample ports and remove all soot and/or moisture from the lines prior to attaching the sample tubing.
-
The analyzer must be stored at ambient temperature (65-80°F) for at least four hours prior to operation.
-
When the unit is to be returned to SSI for service or any other reason, it should be shipped in its original protective packaging. If this packaging is not available, protect the instrument with at least four inches of foam or other impact-absorbing material.
-
For maximum battery life, do not charge the unit until the “battery low” indicator on the display is illuminated.
-
This unit is not designed to measure the dew points in corrosive gases, such as ammonia (NH3), sulphur trioxide (SO3), chlorine (Cl), and hydrochloric acid (HCl).
-
Please read and understand this Operations Manual before operating the unit.
Failure to comply with these conditions may cause damage to the unit that will not be covered under the warranty. Super Systems, Inc. is not responsible for damage to this unit caused by disregard of these warnings, neglect, or misuse.